French English

KermaX plus SDP

Description

KermaX plus SDP

Compact measuring system consisting of a rectangular, transparent ionization chamber integrated electronics and a 10digit internal background lighting LCDisplay, with test / reset and print buttons; optional RS 232 / RS 485 interface for connecting a printer or a PC (RIS / HIS) system. Our quick and easy solution for costeffective retrofit installations at overcouch tube systems.

Specifications

  • Reproducibility: < 1 % (at a constant pressure and temperature)
  • Energy dependence: Better than ± 8 % related to 100 kV according to IEC 60580 from 40 to 150 kV
  • Dose Area Product Rate: Minimum: 0,01 μGym 2 /s Maximum: 3 000,00 μGym² /s
  • Maximal Measurable Dose Area Product: 99999999,99 μGym²
  • Minimal Dose Resolution: 0,01 μGym² (HS High Sensitive, as requested in IEC 60580 table 6)
  • Linearity: Better than ± 2 %
  • Active Area: 140 mm x 140 mm
  • Optical Transperency: ≥ 75 %
  • Voltage Range: 15-28V DC
  • Cabling: Cost effective low voltage cable based on telecommunication standards or network patch cable
  • Mechanical Adaption: Can be mounted directly on the collimator by using the appropriate adaptor rails (distance 176 mm or 167 mm, others on Request)
  • Measuring Time: Minimum 0.01 s
  • Attenuation Equivalent: < 0.5 mm Al
  • Electrode Spacing: 4 mm
  • Chamber Voltage: 410 V ± 5 %
  • AC / DC Converter: 95.9 mm x 55.5 mm x 30.5 mm (L x W x H)
    Not suitable for installing in the collimator.
    Connection cable for KermaX plus: 650 mm long
    Connection cable (input voltage): 900 mm long
    Maximum current consumption: 1 A
    Input Voltage:
    20 50V DC
    14 35V AC
    When Using the AC / DC Converter it is absolutely essential that international standards and national laws are observed. NOT to be installed directly into the Xray System.

Models

  • 120-TinO-IDP - KermaX plus TinO IDP
  • 120-TinO-DDP - KermaX plus TinO DDP
  • 120-IDP - KermaX plus IDP
  • 120-SDP - KermaX plus SDP
  • 120-DDP D - KermaX plus DDP "Duo"